19052 Hipot Tester
|
Chroma19052 |
Instek GPI-735 |
Extech 7410
Associated Research 7650 |
Hioki 3159 |
Kikusui TOS8870A |
Withstanding Voltage Test |
Output voltage |
AC: 0.05-5KV
DC: 0.05-6KV |
AC: 0.1-5KV
DC: 0.1-6KV |
AC: 0-5KV
DC: 0-6KV |
AC:
0-2.5 kV/
0-5kV |
AC:
0-2.5 kV/
0-5kV |
Load Regulation |
±(1%+5V) |
1%+5V |
±(1% of output+5V) |
±1.5% of reading |
Better than 20% |
Voltage Resolution |
2V |
5V |
1V |
Manual adjustment |
NA |
Voltage Accuracy |
±(1%+5counts) |
±(1%of reading + 10V) |
±(2% of setting+5V) |
±5% of reading |
±1.5% of full Scale |
Cutoff Current |
AC: 30mA
DC: 10mA |
AC: 0.1-40mA
DC: 0.1-10mA |
AC: 30mA
DC: 10mA |
AC: 0.01-120mA
DC: - |
AC: 0.5/1/2/4/8/
10/100mA |
Current Resolution |
AC: 1uA
DC: 0.1uA |
AC: 10uA
DC: 10uA |
AC: 1uA
DC: 0.1uA |
AC: 20uA
DC: - |
- |
Current Accuracy |
±(1%+5counts) |
±(1% reading +50uA) |
±(2% f.S+2counts) |
±3% of reading |
±5% of high limit
±20% of low limit |
Current Frequency |
50Hz /60Hz |
50Hz / 60Hz |
50Hz /60Hz |
SYNC AC Frequency |
50Hz /60Hz |
Test time |
0.1-999sec,con |
0.5-999.9sec |
0,0.3-999.9sec |
0.5-999sec |
1-999sec |
Ramp time |
0.1-99.9sec,off |
0.5-999.9sec |
0.1-999.9sec |
NA |
|
ARC Detection Current |
AC: 1-15/DC: 1-10mA |
AC/DC: 1-40mA |
AC/DC: 1-9Range |
- |
- |
Insulation Resistance Test |
IR Range |
0.1M-50GW |
1MW-10GW |
1M-10GW |
1M-2GW |
1M-2GW |
Output Voltage |
50-1000V |
50/100/500/1000V |
100-1000V |
500V/1000V |
500V/1000V |
Voltage Resolution |
2V |
4 Range |
1V |
(Only Two Range) |
(Only Two Range) |
Voltage Accuracy |
±(1%+5V) |
1%of reading +
10V |
|
|
500V:±5%
1000V:±10% |
Resistance Accuracy |
Typical:
±10%+10Count
10-50G:
±15%+10Count |
1MΩ ~ 50MΩ:
±(5%Reading±1Count)
51MΩ ~ 200MΩ: ±(10%Reading±1Count)
201MΩ ~ 2000MΩ: ±(20%Reading±1Count) |
1-1GW:±(2% of
Setting+2counts
1-10GW:±(8% of
Setting+2counts |
±4% of
reading |
|
OSC function
(patent) |
YES |
- |
- |
- |
- |
Back to top
Go back one page |
19053 & 19035 Hipot Tester
|
Chroma 19053 |
Chroma 19035
|
Extech 7420
Associated Reasearch 7650 |
MicroTest 8820 |
Withstanding Voltage Test |
Output voltage |
AC: 0.05-5KV
DC: 0.05-6KV |
AC: 0.05-5KV
DC: 0.05-6KV |
AC: 0-5KV
DC: 0-6KV |
AC: 0-5KV
DC: 0-6KV |
Load Regulation |
±(1%+5V) |
±(1% setting + 0.1%full Scale) |
±(1% of output+5V) |
- |
Voltage Resolution |
2V |
2V |
1V |
10V |
Voltage Accuracy |
1%+5counts |
1% setting + 0.1%full Scale |
±(2% setting+
5V) |
±5% setting |
Cutoff Current |
AC: 30mA
DC: 10mA |
AC: 30mA
DC: 10mA |
AC: 30mA
DC: 10mA |
AC: 10mA
DC: 5mA |
Current Resolution |
AC: 1uA
DC: 0.1uA |
AC: 1uA
DC: 0.1uA |
AC: 1uA
DC: 0.1uA |
- |
Current Accuracy |
±(1%+5counts) |
±(1% Reading
+ 0.5% Range) |
±(2% of setting+
2counts) |
- |
Current Frequency |
50Hz /60Hz |
50Hz /60Hz |
50Hz /60Hz |
50Hz /60Hz |
Test time |
0.1-999sec,continue |
0.1-999sec, continue |
0.1-999sec, continue |
0.1-∞ |
Ramp up time |
0.1-99.9sec,off |
0.1-99.9sec,off |
0.1-999.9sec |
0.1-99.9sec |
ARC Detection Current |
AC: 1-15mA
DC: 1-10mA |
AC: 1-15mA
DC: 1-10mA |
1-9 Range |
- |
Insulation Resistance Test |
Output Voltage |
50-1000V |
50-5000V |
100-1000V |
100-1000V |
Voltage Resolution |
2V |
2V |
1V |
|
Voltage Accuracy |
1.5%+5V |
1%Set + 0.1%full Scale |
±(2% of setting+
2counts) |
±5% |
IR Range |
1M-10GW |
1M-50GW |
1M-10GW |
1MΩ~10GΩ |
Scanner Unit |
8 ports, ±phase |
8 ports, ±phase |
8 ports, ±phase |
8 ports, ±phase |
OSC function
(patented) |
YES |
YES |
- |
- |
DCR Measure |
- |
Y |
- |
- |
Back to top
Go back one page |
19071 Hipot Tester
|
CHROMA 19071
|
Instek 705 |
Extech 7130
Associated Research 3705 |
Hioki 3158 |
Kikusui TOS5050 |
Withstanding Voltage Test |
Output Voltage |
AC: 0.1-5KV |
AC: 0.1-5KV |
AC:0-5KV |
AC:0-2.5KV
/0-5KV |
AC:0-2.5KV
/0-5KV |
Voltage Resolution |
2V |
5V |
10V |
Manual adjustment |
10V |
Voltage Accuracy |
±(1%+5counts) |
±3%of reading |
±(2%of reading+2count) |
±(1.5% of reading) |
±1.5% of full Scale |
Load Regulation |
±(1%+5V) |
±(1%+5V) |
±(2% of output+5V) |
? |
±5% of full Scale |
Cutoff Current |
AC: 0.1-20mA |
AC: 0.1-40mA |
AC:0-20mA |
AC: 0.01-120mA |
AC: 0.1-110mA |
Current Resolution |
1uA |
10uA |
10uA |
10uA |
0.1-9.9mA:0.1maA
10-110mA:1mA |
Current Accuracy |
1.5%+5counts |
±1% of reading |
±(2% of reading+2count) |
±3% of reading |
±(5%+20uA) of
Upper cutoff |
Current Frequency |
50 / 60Hz |
50 / 60Hz |
50 / 60Hz |
50 / 60Hz |
50 / 60Hz |
Test Time |
0.3-999sec
continue |
1-180sec |
0.1-999sec |
0.5-999sec |
0.5-999sec |
Ramp time |
0.1-999sec,off |
0-180sec,off |
0,0.2-999,9sec |
No |
- |
ARC
Detection Current |
1-15mA |
1-40mA |
- |
- |
- |
OSC function
(patented) |
YES |
- |
- |
- |
- |
Memory |
10 Tests,
10 Steps Each |
6 Tests,
6 Steps Each |
10 Tests,
3 Steps Each |
20 Tests |
No |
CE mark |
Y |
Y |
N |
N |
Y |
Back to top
Go back one page |
19073 Hipot Tester
|
Chroma 19073 |
Instek 735 |
Extech 7142
Associated Research 3770 |
Hioki 3153 |
Kikusui TOS5051 |
Withstanding Voltage Test |
Output voltage |
AC: 0.05-5KV
DC: 0.05-6KV |
AC: 0.1-5KV
DC: 0.1-6KV |
AC: 0-5KV
DC: 0-6KV |
AC: 0.2-5KV
DC: 0.2-5KV |
AC: 0-2.5/0-5KV
DC: 0-2.5/0-5KV |
Load Regulation |
±(1%+5V) |
±(1%+5V) |
±(1% of output+5V) |
AC:7%
DC:16% |
No Load
AC: 15%, DC: 3% |
Voltage Resolution |
2V |
5V |
10V |
10V |
10V |
Voltage Accuracy |
±(1%+5counts) |
1%of reading + 10V |
±(2%of reading+
5V) |
±1.5% of reading
±2dgt |
±1.5% of full
Scale |
Cutoff Current |
AC: 0.1-20mA
DC: 0.01-5mA |
AC: 0.1-40mA
DC: 0.1-10mA |
AC: 0-20mA
DC: 0-7.5mA |
AC:0.01-100mA
DC:0.01-10mA |
AC: 0.1-110mA
DC: 0.1-11mA |
Current Resolution |
AC: 1uA
DC: 0.1uA |
AC: 10uA
DC: 10uA |
AC: 1uA
DC: 0.1uA |
AC: 10uA
DC: 10uA |
AC: 10uA
DC: 10uA |
Current Accuracy |
1.5%+5Counts |
1% reading +50uA |
±(2% of setting+
2Counts) |
±(2% f.S.+5 dgt) |
±(5%+20uA) of
Upper cutoff |
Current Frequency |
50Hz / 60Hz |
50Hz / 60Hz |
50Hz / 60Hz |
50Hz / 60Hz |
50Hz / 60Hz |
Test time |
0.1-999sec,continue |
|
0.1-999sec,con |
0.3-999sec |
0.5-999sec |
Ramp time |
0.1-999sec,off |
|
0.1-999.9sec |
0.1-99.9sec |
NA |
ARC Detection Current |
AC: 1-15mA,DC: 1-5mA |
AC/DC: 1-40mA |
AC/DC: 0, 1-9 Range |
- |
- |
Insulation Resistance Test |
IR Range |
0.1MW-50GW |
1MW-10GW |
1MW-10GW |
0.1MW-10GW |
- |
Output Voltage |
DC: 50-1000V |
DC: 50-1000V |
DC: 30-1000V |
DC: 50-1000V |
|
Voltage Resolution |
2V |
5V |
10V |
10V |
- |
Voltage Accuracy |
5%+5Counts |
1%of reading + 10V |
±(2% of setting+2V) |
±0.2 of reading |
- |
OSC function
(patent) |
YES |
- |
- |
- |
- |
Back to top
Go back one page |
19032 Safety Analyzer |
|
Chroma 19032 |
Chroma 19032P |
Instek 745A
Vitrek V4 |
Extech 7742
Associated Research 8106 / 7704 |
Withstanding Voltage Test |
Output voltage |
AC: 0.05-5KV
DC: 0.05-6KV |
AC: 0.05-5KV
DC: 0.05-6KV |
AC: 0.1-5KV
DC: 0.1-6KV |
AC/DC: 0-5KV |
Load Regulation |
1%+5V |
1%+5V |
1%+5V |
±(2% of output+5V) |
Voltage Resolution |
2V |
2V |
5V |
10V |
Voltage Accuracy |
±(1%+5counts) |
±(1%+5counts) |
±(1% of reading+10V) |
±(2% of reading+2count) |
Cutoff Current |
AC: 40mA
DC: 12mA |
AC: 100mA
DC: 20mA |
AC: 40mA
DC: 10mA |
AC: 0-40mA
DC: 0.3-20mA |
Current Resolution |
AC: 1uA
DC: 0.1uA |
AC: 1uA
DC: 0.1uA |
AC: 20uA
DC: 10uA |
AC: 1uA
DC: 1uA |
Current Accuracy |
±(1%+5counts) |
±(1%+5counts) |
±(1% of reading+ 50uA) |
±(2% of reading+2counts) |
Current Frequency |
50Hz / 60Hz |
50Hz / 60Hz |
50Hz / 60Hz |
50Hz / 60Hz |
Test time |
0.3-999sec,con |
0.3-999sec,con |
1-180sec,con |
0-999.9sec |
Ramp up time |
0.3-99.9sec,off |
0.3-99.9sec,off |
0-180sec,off |
0.1-999.9sec,off |
ARC Detection Current |
AC: 1-40mA
DC: 1-12mA |
AC: 1-40mA
DC: 1-12mA |
40 Level |
1-9 Range |
Insulation Resistance Test |
Output Voltage |
0.05-1KV |
0.05-1KV |
50/100/500/1000V |
50V-1000V |
Voltage Resolution |
2V |
2V |
5V |
1V |
Voltage Accuracy |
±(2%+0.5% Range) |
±(2%+0.5% Range) |
±(2%+5V) |
±(7% of reading+2count) |
IR Range |
1M-50GW |
1M-50GW |
1M-10GW |
0.05M-50GW |
Resistance Resolution |
0.1MW |
0.1MW |
1MW |
AVG 0.1MW |
Resistance Accuracy |
5% typical |
5% typical |
1-500MW: 5% of reading
501M-2GW: 10% of reading
2001M-10GW: 20% of reading |
AVG
(7% of reading+2counts) |
Ground Resistance Test |
Output Current |
AC: 1-30A 60A(option) |
3-40A |
DC: 3-42A |
AC: 1-40A |
Current Resolution |
10mA |
10mA |
10mA |
10mA |
Current Accuracy |
±(1%Read+0.2%Range) |
±(1%Read+0.2%Range) |
±1% of read±50mA |
AVG ±(2.5% of reading+2counts) |
GR Range |
10-510mW |
10-510mW |
0-600mW |
0-600mW |
Resistance Resolution |
0.1mW |
0.1mW |
100uW |
1mW |
Resistance Accuracy |
±(1%Reading+0.1%Range) |
±(1%Reading+0.1%Range) |
±(1% of reading+2mW) |
±(3% of reading+2counts) |
Twin-Port Function |
YES |
YES |
- |
- |
Scanner Unit (option) |
8 ports, ±phase
5HV3GB , 3HV5GB
8HV,4HV4GB |
8 ports, ±phase
5HV3GB , 3HV5GB
8HV,4HV4GB |
8HV+8GB
16HV |
8HV+8GB |
Lc Probe |
300V,20A,
RMS, Peak,P1-P2 |
300V,20A,
RMS, Peak,P1-P2 |
- |
277V,15A
RMS |
19032 Leakage Current Test Comparison
|
Chroma |
Extech |
Leakage Current |
Model |
6000-04 |
6000-07 |
6000-08 |
7740Opt.726 |
7742Opt.739 |
DUT Voltage
(Line input) |
300V |
300V |
300V |
277V |
277V |
DUT Voltage Accuracy |
±(1.5%Reading
+0.2% full scale) |
±(1.5%Reading
+0.2% full scale) |
±(1.5%Reading
+0.2% full scale) |
±(1.5%Reading
+2Counts) |
±(1.5%Reading
+2Counts) |
DUT Current |
10A |
20A |
20A |
15A |
15A |
Current Accuracy |
±(1.5%Reading
+0.5% full scale) |
±(1.5%Reading
+0.5% full scale) |
±(1.5%Reading
+0.5% full scale) |
±(2%Reading
+2Counts) |
±(2%Reading
+2Counts) |
Leakage Current
Measurement |
RMS |
RMS |
RMS+ Peak
AC+DC |
RMS |
RMS+ Peak
AC+DC |
Leakage Current Accuracy(RMS) |
±(2%Reading
+1% full scale) |
±(2%Reading
+1% full scale) |
±(2%Reading
+1% full scale) |
±(1.5%Reading
+3Counts) |
±(1.5%Reading
+3Counts) |
Leakage Current Accuracy(Peak) |
- |
- |
±(5%Reading
+1% full scale) |
- |
±(10%Reading
+2uA) |
P2 to G ,
P1 to P2 |
- |
Y |
Y |
- |
Y |
Back to top
Go back one page |
The information in this document was gathered from readily available data and is subject to change by each manufacturer without notice, for more Information please call us.
|